Titan and Leginon Check List For Automated Collection
- Set defocus range (e.g. 0.8 – 3.0 microns)
- Targets are submitted
- User verification is OFF
- Turbo pump OFF
- Objective in and centered
- Astigmatism
- Rotation center
- Dose match
- New gain references for final exposure time
- Field compensator reset after final exposure adjustments
- Save frames data
- All samples are put away
- Liquid Nitrogen level
Below is a description of the framesdata pipeline.
Caution – under construction.
- Copying frames:
- Start the robocopy script to transfer from FastSSD -> D: drive to Archive -> E: drive
- Open a command prompt on the DE computer
- Enter the following command: robocopy D:\RawFrames\DE20\ E:\Data\Integrating\ *.mrc /s /ts /v /r:2 /w:2 /move /mot:5 /r:2 /w:5 (the script is also in a file on the DE comp desktop called commands.rtf)
- On hermes -> Load the module: myamidev
- Change group to cryo_secm4
- Run the transferFrames2.py script -> transferFrames2.py --origpath /mnt/176_Data/Integrating/ --despath /mnt/obscure_lustre/cryo/lustre/framesdata --session 17jan09a --preset ex
.
.
.